Tibial Preparation
Tibial Sizing
- Determine the tibial implant size using the tibial viewing template.
- Place the appropriate tibial drill guide or stemless tibial trial on the tibia.
- Centralize and pin the tibial drill guide or stemless trial.
- Drill through the stemless trial or place the 11 mm tibial collet in the drill guide and drill with the 11 mm tibial drill.If a 9.5mm drill has been used for the intramedullary tibial alignment assembly, only the 11 mm tibial punch is needed.
- Punch with the 11 mm tibial punch. (Figure 17)
- Remove the tibial drill guide if used and place the tibial trial onto the proximal tibia to assess coverage.
|
|